F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
E

Respiratory Care and Equipment Cleaning Deficiencies

West Anaheim Medical Center D/p SnfAnaheim, California Survey Completed on 03-27-2026

Summary

The facility failed to provide necessary respiratory care services for four residents who were receiving suctioning and ventilator-related care. For Resident 2, the physician’s orders included oxygen therapy via trach collar, suctioning every two hours and as needed, and changing the Yankauer suction device every 24 hours and as needed. During observation, an opened Yankauer suction device was found at the bedside with a label showing it had been opened on 3/22/26, and the respiratory therapist stated it should have been changed after 24 hours once opened. For Residents 3 and 12, both had care plans and physician’s orders directing that the Yankauer suction be changed daily and as needed. During observation, each resident had an opened Yankauer suction tip stored in the original packaging at the bedside, and neither package was labeled with the opened date. The respiratory therapist stated the Yankauer suction tip should be labeled when opened and changed every shift for infection control purposes, and later discarded the suction tips for both residents. The DON reviewed the records and verified the orders and care plans called for changing the Yankauer suction every 24 hours. For Resident 15, who was admitted with a tracheostomy and ventilator support, the ventilator operator’s manual directed that device parts be cleaned with warm water and soap or mild detergent, rinsed thoroughly, and air dried. The resident’s record showed ventilator settings and respiratory treatment documentation, but there was no documented evidence that the ventilator machine had been cleaned and disinfected per the manufacturer’s instructions. During interviews, respiratory staff were uncertain who was responsible for cleaning the ventilator, and one RT stated the facility did not have a specific regular cleaning and disinfecting process for ventilator machines and instead cleaned them with bleach wipes as needed. The DON was informed and verified these findings.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Incorrect Oxygen Flow Rates and Delayed Tubing Changes
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Two residents receiving O2 had flow rates set above their physician orders, and one resident’s tubing remained dated beyond the weekly change interval. Staff, including an RN and the DON, confirmed the mismatched flow rates and noted that tubing should be changed weekly by the RT. The facility policy required O2 to be administered per physician order and tubing/cannula to be changed weekly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Concentrator Left in Room After Order Discontinued
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Oxygen Concentrator Left in Room After Order Discontinued: A resident with morbid obesity with alveolar hypoventilation and diabetes was observed with an oxygen concentrator at bedside even though there was no current oxygen order. The resident stated he used oxygen only when needed, and the CNO confirmed the oxygen had been discontinued earlier and the concentrator should not have remained in the room.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Order for CPAP Use and Maintenance
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Missing Order for CPAP Use and Maintenance: A resident with sleep apnea and stroke had a CPAP in the room, but there was no physician order or care plan documentation for its use or maintenance. Staff knew the resident had the CPAP, yet CNAs and an RN reported they did not routinely clean the mask or tubing, and the LPN Resident Care Manager confirmed no order existed for the CPAP or its care.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Maintained or Applied as Ordered
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with a continuous O2 order was found without oxygen in place, and an LPN later said the order had been misread. Two other residents with COPD had respiratory equipment that was dirty or outdated, including a gray concentrator filter, tubing and water bottles dated weeks earlier, and tubing lying on the floor with visible dust. Staff gave inconsistent accounts of who was responsible for changing or cleaning the equipment, and the DON acknowledged the dirty equipment and old mask placed residents at risk for respiratory infection or pneumonia.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Not Properly Delivered via Nasal Cannula
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with acute respiratory failure with hypoxia and severe cognitive impairment was ordered oxygen via NC at 2 to 5 LPM to keep O2 saturation at or above 95%. During observation, the resident was lying in bed with the NC on the side of the face instead of in the nose while the concentrator was set at 2.5 LPM. RN and DON stated the NC should be in the nose to deliver oxygen as ordered.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Incorrect Oxygen Flow Settings
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD and chronic respiratory failure did not receive oxygen at the ordered flow rate. Staff found the oxygen concentrator set below the ordered amount on multiple observations, and the resident stated the oxygen was supposed to be set higher. The DON, LVNs, and ADON acknowledged the setting was incorrect and that the ordered flow should have been followed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across California

Get a heads-up on the newest immediate-jeopardy (J–L) citations in California — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.