F0908 F908: Keep all essential equipment working safely.
F

Kitchen Drainage Air Gaps Not Maintained

Anberry Nursing And Rehabilitation CenterAtwater, California Survey Completed on 01-23-2026

Summary

Essential kitchen equipment was not kept in safe operating condition when the food preparation sink did not have an air gap and the ice machine that supplied ice to residents also did not have an air gap. During a concurrent observation and interview with the Dietary Manager in the kitchen, the food prep sink was seen piped directly into the wall, and a drain on the floor beneath it had no piping going to it. The Dietary Manager stated the facility was not using an air gap and that the pipe from the food prep table went directly into the city sewer. During a concurrent observation and interview in the breakroom, the ice machine was observed with the hose hanging over the rim of the drain and passing below the level of the floor. The Dietary Manager stated this was the only ice machine in the facility and that there was no air gap for it. The Maintenance Director and Registered Dietician both stated the food prep sink and ice machine lacked the required air gaps, and the facility policy titled Drainage Air Gap Policy stated air gaps were required for drains in all areas of the facility, including ice machines, dish machines, and floor drains under sinks.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0908 citations
Expired Supplies Found in Crash Carts
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Expired supplies were found in two crash carts on nursing units 1B and 2B. An opened Ambu bag in the 2B cart was expired, and the 1B cart contained two expired Ambu bags and expired Assure Prism glucometer strips. RN staff confirmed the items were expired and should not have been in the carts.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Kitchen Stove Burner Not Functioning
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

A gas stove in the kitchen had a back center burner that would not ignite when turned on, and staff reported it had been not working for about 1 to 2 months. Interviews showed cooks were expected to report equipment problems to the DM, who would notify Maintenance, but the issue was not promptly communicated to Maintenance. No gas odor or hissing was observed, and the burner was later taken out of service until repaired.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Elevator Malfunctioned and Remained in Need of Repair
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

The facility failed to keep one of its elevators in safe operating condition. Records showed elevator two malfunctioned multiple times and needed a replacement part, and residents and staff were aware of the repeated malfunctions. The Administrator confirmed the elevator was malfunctioning and needed repair.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Dryer Lint Traps
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Failure to Maintain Dryer Lint Traps: Surveyors observed heavy lint buildup in both dryer lint traps, with lint hanging from each trap and covering the full trap surface. The sign-off log showed no entries after the prior shift, and staff interviews confirmed the required two-hour checks and documentation were not consistently completed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Glucometer Calibration and Quality Control Not Documented
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Glucometer calibration and QC were not documented for one glucometer on Medication Cart D, and the QC log was blank for the scheduled check. LVN and DON verified that the 11-7 shift was responsible for nightly/daily calibration and QC to support accurate blood sugar readings.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Thermostats Not Powered On or Working
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Thermostats Not Powered On or Working: Two of five sampled thermostats were observed turned off with black screens and no visible power. A resident’s RP said it was always hot in the resident’s room and a fan was provided because there was no working AC. The MS stated the thermostats should be on and working to regulate room temperatures, but he was unsure how often they should be checked.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across California

Get a heads-up on the newest immediate-jeopardy (J–L) citations in California — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release October 8, 2026) and official state health department websites — never guesswork.