Glucometer Quality Control Logs Contained Out-of-Range and Illegible Entries
Summary
The facility failed to ensure 4 of 4 glucometers used for residents with orders for blood glucose monitoring were maintained in safe operating condition. The facility’s Glucometer Information Sheet stated that the EvenCare 2 glucometer was in use, that control testing was to be completed at least daily, and that any out-of-range control reading would not be used for resident testing until the glucometer was repaired or repeat control tests were within normal limits. The Director of Nursing reported the facility had 29 residents who required blood glucose monitoring with the facility’s glucometer. Review of the November 2025 and December 2025 Glucometer Quality Control Logs showed multiple control results documented as within range when the recorded values were outside the acceptable ranges for glucometers #1, #2, #3, and #4. Some entries were illegible or overwritten, including a high control reading that appeared to be 232 and another recorded as 1, and several out-of-range readings had no comments explaining how they were handled. For glucometer #1 in November 2025, the log also failed to include lot numbers, expiration dates, and acceptable ranges for the low and high control solutions. During interview, the DON reviewed the logs and confirmed the illegible control values, out-of-range control values without comments, and the missing documentation for glucometer #1.
Penalty
Resources
Below are regulatory guidelines relevant to this citation:
Trusted data from CMS and state health departments
Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release October 8, 2026) and official state health department websites — never guesswork.
In your survey window? See what surveyors are citing.
The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.