F0908 F908: Keep all essential equipment working safely.
E

Failure to Maintain Functioning Nurse Call System

Pruitthealth-north Tampa, LlcLutz, Florida Survey Completed on 10-18-2024

Summary

The facility failed to maintain a functioning nurse call system, which was observed to be non-operational during a two-day survey. Residents reported that their call lights were not working, and they were provided with table bells as an alternative, which were also ineffective in summoning assistance. Resident #9 and Resident #8's representative both confirmed the call lights were not functioning, and the table bells provided did not result in timely assistance. The Director of Nursing initially claimed all systems were functional, but later acknowledged awareness of the issue in some rooms. The Director of Maintenance (DOM) admitted to being aware of the malfunctioning call lights in certain rooms and had issued table bells as a temporary solution. However, a complete facility audit was not conducted, and only a sampling of rooms was checked weekly. The DOM demonstrated altering the logbook documentation, changing a 'Fail' to 'Pass', which was claimed to be an error. Staff members, including a Registered Nurse, reported the issue to administration, but not all affected residents received table bells due to a shortage. The Nursing Home Administrator (NHA) stated that a vendor was contacted for repairs, but the facility was busy due to recent hurricanes. The NHA believed that providing table bells met the requirement for residents to notify staff of their needs. However, the facility lacked a policy and procedure for Equipment Repair and Maintenance, contributing to the ongoing issue with the nurse call system.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0908 citations
Expired Supplies Found in Crash Carts
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Expired supplies were found in two crash carts on nursing units 1B and 2B. An opened Ambu bag in the 2B cart was expired, and the 1B cart contained two expired Ambu bags and expired Assure Prism glucometer strips. RN staff confirmed the items were expired and should not have been in the carts.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Kitchen Stove Burner Not Functioning
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

A gas stove in the kitchen had a back center burner that would not ignite when turned on, and staff reported it had been not working for about 1 to 2 months. Interviews showed cooks were expected to report equipment problems to the DM, who would notify Maintenance, but the issue was not promptly communicated to Maintenance. No gas odor or hissing was observed, and the burner was later taken out of service until repaired.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Elevator Malfunctioned and Remained in Need of Repair
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

The facility failed to keep one of its elevators in safe operating condition. Records showed elevator two malfunctioned multiple times and needed a replacement part, and residents and staff were aware of the repeated malfunctions. The Administrator confirmed the elevator was malfunctioning and needed repair.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Dryer Lint Traps
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Failure to Maintain Dryer Lint Traps: Surveyors observed heavy lint buildup in both dryer lint traps, with lint hanging from each trap and covering the full trap surface. The sign-off log showed no entries after the prior shift, and staff interviews confirmed the required two-hour checks and documentation were not consistently completed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Glucometer Calibration and Quality Control Not Documented
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Glucometer calibration and QC were not documented for one glucometer on Medication Cart D, and the QC log was blank for the scheduled check. LVN and DON verified that the 11-7 shift was responsible for nightly/daily calibration and QC to support accurate blood sugar readings.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Thermostats Not Powered On or Working
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Thermostats Not Powered On or Working: Two of five sampled thermostats were observed turned off with black screens and no visible power. A resident’s RP said it was always hot in the resident’s room and a fan was provided because there was no working AC. The MS stated the thermostats should be on and working to regulate room temperatures, but he was unsure how often they should be checked.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Florida

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Florida — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release October 8, 2026) and official state health department websites — never guesswork.