F0908 F908: Keep all essential equipment working safely.
D

Shower Gurney Collapse Due to Missing Pin

The Suites Rio VistaRio Rancho, New Mexico Survey Completed on 07-10-2024

Summary

The facility failed to ensure that patient care equipment was in safe operating condition, resulting in an incident involving a resident with quadriplegia and a traumatic brain injury. The resident was being transferred from his room to the shower room on a shower gurney when the gurney collapsed. This occurred because a pin that supports the head of the gurney was missing, causing the head to fall and the resident to slip from the gurney to the ground. The incident was confirmed through interviews with the LPN and CNA involved, as well as the resident, who recalled the fall and stated that he was not injured but was frightened by the event. The CNA demonstrated that the shower gurney was made of heavy tubes with a thick foam mattress, and the head of the gurney could be released by removing a pin. At the time of the accident, the pin was missing, leading to the collapse. The Assistant Director of Nursing and the Administrator confirmed the equipment failure and acknowledged that the gurney had since been repaired. They also confirmed that staff had been educated to check the equipment before use to ensure all parts are in place and secure.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0908 citations
Inoperable Commercial Washer in Laundry Department
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

In the laundry department, 1 of 2 commercial washers was out of service for more than 8 months, leaving only 1 washer available for resident laundry. Laundry Aides stated they used the working washer for all residents and that keeping up with timely laundry services was sometimes challenging with only 1 machine.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Hoyer Lift Batteries Not Maintained in Safe Operating Condition
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

A resident who required a mechanical lift for transfers experienced repeated Hoyer lift battery failures during transfers, including one observed transfer where the lift stopped working while he was being lowered. Staff reported that lift batteries were often not charged, that overnight staff were responsible for charging them, and that they sometimes used the emergency release to lower the resident when the battery died. The DON and other staff confirmed the batteries were not consistently checked or maintained, and the maintenance supervisor said batteries were only replaced occasionally.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Flooring Renovation Started Before Construction Review Approval
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

Flooring Renovation Started Before Construction Review Approval: The facility removed carpet and installed laminate flooring in resident areas before receiving DOH Construction Review Services approval. The Administrator said the second-floor west unit remodeling had already been completed and residents were moved to another unit during the work. Notices to residents described planned flooring replacement, observation confirmed new wood flooring in the hallway and resident rooms, and a resident said they were relocated for at least two weeks. The Regional Plant Operations Director stated the approval was still pending and that they were unsure of the approval process.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Laundry Equipment Not Maintained in Safe Operating Condition
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

Laundry equipment was not maintained in safe operating condition. A resident with ESRD on dialysis reported that clothes came back stinking, while the Laundry Supervisor said the industrial washer kept breaking down, hot water was not getting hot, the wrong soap was being used for resident clothing, and the department lacked a hopper for washing soiled linens. Observation showed a small washer overfilled with sheets, and staff gave mixed reports about washer breakdowns and linen shortages.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Mechanical Lift Found With Exposed Charging Cord
D
F0908 F908: Keep all essential equipment working safely.
Short Summary

Mechanical lift equipment on the 3rd floor was observed with exposed black and red inner cords from the grey charging cord hanging out. An RN said the lift should not be used if the cords are exposed, and the DON stated such equipment should be reported to maintenance and removed from the floor for safety precautions. The Maintenance Director confirmed the outer grey cord protects the inner cords, and the maintenance log showed no repair report for the lift.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Expired Ambu Bags Found on Two Crash Carts
E
F0908 F908: Keep all essential equipment working safely.
Short Summary

Expired Ambu Bags Found on Two Crash Carts: The facility failed to ensure two crash carts were in safe operating condition when an Ambu bag on the First Floor Crash Cart and an Ambu bag on the Second Floor Crash Cart were both found expired. RN staff confirmed the expired equipment and acknowledged the carts were not maintained as required by the facility's crash cart management process.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across New Mexico

Get a heads-up on the newest immediate-jeopardy (J–L) citations in New Mexico — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments