F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
E

Oxygen Therapy Deficiencies

Oasis Nursing & Rehabilitation CenterEl Paso, Texas Survey Completed on 06-25-2026

Summary

The facility failed to provide safe and appropriate respiratory care for two residents receiving oxygen therapy. For Resident #32, who had COPD, chronic respiratory failure with hypoxia, pan lobular emphysema, and a BIMS score of 7, the record showed that oxygen therapy was part of the care plan and that staff were to administer oxygen as ordered, monitor respiratory status, and maintain oxygen equipment in safe working condition. During an observation, the resident was seen wearing a nasal cannula and receiving oxygen while eating breakfast, but there was no oxygen warning sign posted outside the room at that time. Interviews with an LVN, another LVN, a CNA, the DON, and the Administrator confirmed that oxygen-in-use signs were required outside rooms of residents receiving oxygen therapy and were intended to alert staff, visitors, and others that oxygen was present. They stated that the signs were needed so appropriate safety precautions could be followed and that the absence of the sign could allow staff or visitors to enter without knowing oxygen was in use. The facility policy titled Oxygen Administration stated that no smoking signs were to be placed when oxygen was administered and stored, and also stated that oxygen-in-use signs were not required on individual resident rooms if the facility was non-smoking. For Resident #12, who had ESRD, dependence on renal dialysis, cognitive communication deficit, unspecified dementia without behavioral disturbance, and a BIMS score of 7, the record review showed no documentation in the care plan addressing oxygen concentrator filter maintenance. During observation, the resident’s oxygen concentrator air filter was noted to be dirty and covered with dust and lint. CNA A, RN F, the ADON, the DON, and the Administrator all acknowledged that the filter was dirty. The ADON stated the facility did not have a specific policy for cleaning or maintaining oxygen concentrator filters, staff were told to follow manufacturer recommendations, and there was no documentation showing when filters were cleaned or who cleaned them. The DON stated that central supply was responsible for cleaning filters and nursing staff were responsible for cleaning the concentrators, and that the facility had not provided in-service education regarding oxygen concentrator filter maintenance.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Oxygen equipment not maintained per order
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident who required oxygen therapy for SOB had an oxygen bubbler that was overdue for replacement, despite a TAR directing weekly changes of oxygen supplies including the bubbler and tubing. Staff confirmed the bubbler should have been changed weekly, and the DON and staff educator stated nurses were expected to follow the provider order as written and that all disposable oxygen pieces were to be changed weekly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Bi-pap Therapy Not Ordered or Verified at Admission
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD, acute respiratory failure with hypercapnia, and oxygen dependence was admitted with prior records showing nightly Bi-pap use and specific settings, but the facility did not have a Bi-pap order or settings in the chart and the baseline care plan did not address oxygen or Bi-pap. The resident reported he went without Bi-pap for three nights because a connector piece was missing, while the hospice RN, admitting LPN, ADON, and DON each stated they were unaware of the machine or did not verify the equipment and orders at admission.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Care and Tracheostomy Care Not Provided as Ordered
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory care was not provided as ordered for one resident receiving O2, as the NC tubing was not dated. Tracheostomy care was also not provided as ordered for another resident: heavy thick mucus was visible under the trach, the ATM mask was soiled with mucus, the ATM and nebulizer tubing were not dated or stored in a bag, and the suction catheter tubing end was not stored in a bag. An RN confirmed the findings.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Change Oxygen Equipment per Physician Orders
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

The facility failed to follow physician orders for oxygen equipment for two residents with COPD-related diagnoses. Observations showed each resident’s O2 tubing and water bottle remained dated and unchanged beyond the ordered weekly schedule, and the DON confirmed the equipment should have been changed per the orders.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Ordered Continuous Oxygen
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Failure to maintain ordered continuous O2: A resident with stroke and COPD was observed without his NC while a CNA waited for a portable O2 unit to be filled. The CNA then applied the NC but did not turn on the liquid portable O2, and later wheeled the resident to the nurse’s station to have the nurse set the unit to the ordered liter flow. The resident’s physician had ordered continuous O2 2L to 4L NC.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage of CPAP Mask and Oxygen Tubing
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Improper Storage of CPAP Mask and Oxygen Tubing: A resident with COPD, DM, dialysis dependence, and continuous O2 orders had his CPAP/NIV mask observed lying on the bedside table and his O2 tubing wrapped under the concentrator handle instead of being stored in a sanitary container or bag. Staff interviews confirmed the equipment should be stored in a bag when not in use, but it was not being kept that way during observation.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Texas

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Texas — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release October 8, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.