F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
D

Missing Bed Rail Maintenance Inspections

Arcadia Care AledoAledo, Illinois Survey Completed on 05-13-2026

Summary

Failed to conduct regular maintenance inspections as part of a maintenance program to identify possible bed entrapment areas for three residents reviewed for side rail use. The facility policy dated 12/25 states that bed rails and mattresses must be properly installed and maintained, and that the mattress and bed rails should be inspected regularly for areas of possible entrapment. The Maintenance Director job description dated 5/26 states the Maintenance Director is responsible for maintaining the facility in a safe and comfortable manner and making weekly inspections of maintenance functions. On 5/11/26, R21 was observed lying in bed with a half side rail raised on the left side; the bed was positioned against the left wall, and R21 was confused with mumbled speech. R21's EHR from 5/11/25 through 5/12/26 did not include a maintenance assessment or evaluation of the side rails. On 5/12/26, R35 was observed lying in bed with a half side rail raised on the left side and the bed positioned against the left wall; R35's EHR from admission on 4/25/25 through 5/12/26 did not include a maintenance assessment or evaluation of the side rails. R40's EHR from admission on 3/5/26 through 5/11/26 also did not include a maintenance assessment or evaluation of the side rails. On 5/11/26, R40 was observed with a half side rail raised on the left side, and another half side rail was lying on the floor behind the head of the bed; R40 stated the side rail on the floor had broken off and no one had fixed it. On 5/12/26, the Maintenance Director stated he had not performed maintenance inspections or assessments for any residents' side rails and could not find any side rail inspections or assessments done within the last year, and he had not been told that R40's side rail had broken off.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0909 citations
Torn Mattress Left in Resident Room
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Torn Mattress Left in Resident Room: A resident’s bed mattress was observed with punctures exposing internal materials and with dips and lumps on the sleeping surface while the bed was unmade. The resident said staff had known about the mattress for over a year and that it was very uncomfortable. The HSKP supervisor said staff usually report worn mattresses, and the administrator said new mattresses had been ordered but was unsure how this one was missed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Inspect Bed Systems for Entrapment Hazards
F
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Failure to inspect bed systems for entrapment hazards: The facility did not maintain a regular program to inspect bed rails, mattresses, and frames or measure FDA entrapment zones. The ML stated annual bed checks focused on electrical components and function, but entrapment measurements were not being done, and the DON stated all residents had bed rails because they came on the bed. Multiple residents were observed with raised bed rails, including combinations of upper and lower rails, and the facility policy called for regular bed inspections and individual bed rail evaluations.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Unsafe Bed and Mattress Inspection and Attachment
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

A facility failed to ensure safe inspection and fit of resident beds and air mattresses. One resident with multiple chronic conditions had an air mattress with an approximately five-inch gap between the mattress and foot board on repeated observations, and staff were unclear who was responsible for correcting it. Another resident fell from bed when a broken clip securing the air mattress strap to the frame allowed the bed to lift; surveyors later observed the mattress was too wide for the bedframe, with about five inches hanging over the side and not fully supported.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Bed Rail Gap Measurements Exceeded Allowed Standards
E
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Bed rail safety checks showed Zone 1 gaps of 7 inches for three residents, exceeding the FDA maximum of 4 3/4 inches. One resident had Bell’s palsy, falls history, reduced mobility, and cognitive impairment; another had Alzheimer’s disease, dementia, a tibia fracture, and needed max assist for bed mobility; and a third had quadriplegia and needed max assist for bed mobility. Observations found large square openings in the rails, a mattress separated from the frame, and staff interviews showed the Housekeeping Supervisor measured the rails while nursing leadership kept the forms and was responsible for compliance.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Lack of Routine Bed Entrapment Inspections
F
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

The facility failed to maintain a routine process to inspect bed frames, mattresses, and bed rails for possible entrapment. Record review showed the bed inspection form addressed bed height and wheelchair cushion securement, but did not include checks of frames, mattresses, or rails. The ADMIN and MAINT both confirmed there was no checklist or routine maintenance program for these inspections unless a work order was submitted.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Bed Rail Inspection Documentation
E
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Missing Bed Rail Inspection Documentation: The facility failed to complete and/or maintain routine bed rail inspection records for all residents with side rails. A resident with bilateral half rails used for repositioning and safety due to frequent falls out of bed was observed with upper quarter-length rails, and maintenance staff from a sister facility later checked the rails without prior facility documentation. An undated bed entrapment inspection form showed 10 beds passed, but it did not identify who completed the assessments.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Illinois

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Illinois — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.