F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
D

Nebulizer Components Left Wet and Uncovered After Treatments

Avera Rosebud Country Care CenterGregory, South Dakota Survey Completed on 05-21-2026

Summary

Infection control practices were not followed for the cleaning of nebulizers for two residents who received scheduled nebulizer treatments. Resident 21, who had diagnoses of shortness of breath, wheezing, and obstructive sleep apnea and a BIMS score of 15, was observed with a nebulizer machine on her nightstand with the tubing and mouthpiece connected and the medication reservoir appearing wet; the mouthpiece was lying directly on the nightstand, uncovered. She stated she received three nebulizer treatments a day and that staff sometimes disconnected the nebulizer to clean it, and that staff usually took the medication reservoir to the bathroom and rinsed it out before giving her treatment. Resident 8, who had COPD and a BIMS score of 9, was observed with a nebulizer machine and attached mouthpiece lying on his bedside table, and the medication reservoir had multiple small wet spots and a mist film inside. He stated he received two nebulizer treatments a day, and staff cleaned the nebulizer components but he was unsure how often they were cleaned during the day. RN C stated she had forgotten to return to resident 8's room to clean the nebulizer components after his morning treatment, and that after each use the mouthpiece, extender, and medication reservoir should be cleaned. The provider's updated May 2026 nebulizer policy stated the components should be rinsed after each treatment, placed on a clean towel and covered with a clean paper towel, and washed with soap and water every 24 hours.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Missing Oxygen Order and Improper Nebulizer Storage
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Missing Oxygen Order and Improper Nebulizer Storage: A resident receiving oxygen for sleep apnea had no physician order specifying the oxygen delivery rate, even though staff were setting the rate. In a separate observation, a resident’s nebulizer mask was left unbagged on the nightstand while not in use. An LPN and the DON both stated respiratory equipment should be stored in a bag when not in use, and the facility policy required bagged storage between uses.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory equipment was not maintained and stored per standards
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory care was not provided in accordance with standards for several residents receiving oxygen and nebulizer treatments. Staff observed outdated oxygen tubing and humidification equipment, a nebulizer mouthpiece left out with tubing touching the floor, and oxygen supplies not stored properly. One resident’s oxygen was running at a higher flow than ordered, and staff confirmed the tubing and humidification items should be changed weekly and documented.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage and Dating of Oxygen and CPAP Equipment
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Improper Storage and Dating of Oxygen and CPAP Equipment: Two residents had oxygen equipment and CPAP items observed out of proper storage, including nasal cannula tubing on the floor, undated or empty humidifiers, and a CPAP mask on the floor. One resident had OSA with an order for oxygen at bedtime and as needed, and the other had COPD with orders for nightly CPAP and continuous oxygen. Staff stated the tubing, humidifier, and CPAP mask should be stored and dated per facility practice, and the facility policy required weekly changes and dating of oxygen equipment.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Contaminated oxygen tubing was placed back on a resident
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident receiving continuous O2 via nasal cannula had the cannula found on the floor while the concentrator was running. A CNA placed the cannula back on the resident’s face before an LVN identified it as contaminated and replaced the cannula and tubing. The resident had diagnoses including hypertensive heart disease and atherosclerotic heart disease, and the care plan called for O2 at 2 L/min to maintain O2 sats above 92%.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Stored Sanitarily
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD, a respiratory infection, continuous O2, and CPAP therapy had CPAP and nasal cannula tubing observed lying on the bed and the CPAP mask draped over the machine instead of being stored in a sanitary container. Staff interviews showed inconsistent understanding of how to store the respiratory equipment, and the facility did not provide the requested storage of the equipment when not in use.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Properly Labeled or Stored
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory equipment was not properly changed, labeled, or stored for a resident receiving nebulizer treatment. An undated aerosol mask attached to a nebulizer was observed on the resident’s nightstand, open to air, and remained there on a later observation. An LPN confirmed the mask was open to air and not stored properly, and the DON confirmed the mask and tubing had not been dated or stored properly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across South Dakota

Get a heads-up on the newest immediate-jeopardy (J–L) citations in South Dakota — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release July 29, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙