F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
D

Unsafe Bed Frame, Mattress, and Bed Rail Conditions

Countryside Manor Nursing And Rehabilitation LlcFremont, Ohio Survey Completed on 05-19-2026

Summary

The facility failed to ensure resident beds were properly maintained in a safe manner for two residents. One resident had diagnoses including congestive heart failure, obesity, chronic respiratory failure, venous insufficiency, depression, atrial fibrillation, type 2 diabetes mellitus, hypertension, and venous thrombosis and embolism, and was dependent on staff for activities of daily living including bed mobility. Although the resident’s care plan included quarter upper bilateral side rails for bed mobility and transfer ability, observation showed the mattress was undersized, leaving approximately four inches of exposed metal bed frame on the right side and a four-inch gap between the mattress and the right upper quarter side rail. When staff turned the resident, the resident’s right foot contacted the metal bed frame with sharp edges, and the LPN verified the mattress was undersized and exposed the frame with a gap between the rail and mattress. The Maintenance Director later measured gaps of six inches at the foot of the bed, four inches in the middle, and three inches between the rail and mattress, and stated he was unaware when the bed and mattress were assessed for proper fit. A second resident had diagnoses including dementia, transient ischemic attack, cerebral infarction, spinal lumbar stenosis, anxiety disorder, paranoid schizophrenia, major depression, intellectual disability, congestive heart failure, and hypertension, and was dependent on staff for bed mobility and transfers with a history of falls. The resident’s care plan and physician order included upper quarter bilateral side rails for bed mobility and transfer ability enabler. Observation showed the bed had a loose quarter side rail on the left side and no quarter side rail on the right side. CNAs verified the left rail was loose and the right rail was missing, and stated the resident used the side rails when being positioned in bed and was fearful of falling without them. Later, an RN also verified the left bed rail was loose and the right bed rail was missing, and the Maintenance Director stated he was unaware of the loose and missing bed rail.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0909 citations
Torn Mattress Left in Resident Room
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Torn Mattress Left in Resident Room: A resident’s bed mattress was observed with punctures exposing internal materials and with dips and lumps on the sleeping surface while the bed was unmade. The resident said staff had known about the mattress for over a year and that it was very uncomfortable. The HSKP supervisor said staff usually report worn mattresses, and the administrator said new mattresses had been ordered but was unsure how this one was missed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Inspect Bed Systems for Entrapment Hazards
F
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Failure to inspect bed systems for entrapment hazards: The facility did not maintain a regular program to inspect bed rails, mattresses, and frames or measure FDA entrapment zones. The ML stated annual bed checks focused on electrical components and function, but entrapment measurements were not being done, and the DON stated all residents had bed rails because they came on the bed. Multiple residents were observed with raised bed rails, including combinations of upper and lower rails, and the facility policy called for regular bed inspections and individual bed rail evaluations.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Unsafe Bed and Mattress Inspection and Attachment
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

A facility failed to ensure safe inspection and fit of resident beds and air mattresses. One resident with multiple chronic conditions had an air mattress with an approximately five-inch gap between the mattress and foot board on repeated observations, and staff were unclear who was responsible for correcting it. Another resident fell from bed when a broken clip securing the air mattress strap to the frame allowed the bed to lift; surveyors later observed the mattress was too wide for the bedframe, with about five inches hanging over the side and not fully supported.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Bed Rail Gap Measurements Exceeded Allowed Standards
E
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Bed rail safety checks showed Zone 1 gaps of 7 inches for three residents, exceeding the FDA maximum of 4 3/4 inches. One resident had Bell’s palsy, falls history, reduced mobility, and cognitive impairment; another had Alzheimer’s disease, dementia, a tibia fracture, and needed max assist for bed mobility; and a third had quadriplegia and needed max assist for bed mobility. Observations found large square openings in the rails, a mattress separated from the frame, and staff interviews showed the Housekeeping Supervisor measured the rails while nursing leadership kept the forms and was responsible for compliance.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Lack of Routine Bed Entrapment Inspections
F
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

The facility failed to maintain a routine process to inspect bed frames, mattresses, and bed rails for possible entrapment. Record review showed the bed inspection form addressed bed height and wheelchair cushion securement, but did not include checks of frames, mattresses, or rails. The ADMIN and MAINT both confirmed there was no checklist or routine maintenance program for these inspections unless a work order was submitted.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Bed Rail Inspection Documentation
E
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Missing Bed Rail Inspection Documentation: The facility failed to complete and/or maintain routine bed rail inspection records for all residents with side rails. A resident with bilateral half rails used for repositioning and safety due to frequent falls out of bed was observed with upper quarter-length rails, and maintenance staff from a sister facility later checked the rails without prior facility documentation. An undated bed entrapment inspection form showed 10 beds passed, but it did not identify who completed the assessments.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Ohio

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Ohio — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.