F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
E

Respiratory Care and Oxygen Therapy Deficiencies

Huntington Valley Healthcare CenterHuntington Beach, California Survey Completed on 05-21-2026

Summary

The facility failed to provide safe and appropriate respiratory care for multiple residents receiving oxygen therapy. During the initial tour and subsequent record review, surveyors observed that Resident 2’s nasal cannula tubing was unlabeled and undated while the resident was receiving oxygen at 2 LPM, and Resident 5’s oxygen tubing was also unlabeled and undated, coiled, and hanging on the oxygen machine rather than being inside a bag when not in use. The facility’s records showed both residents had physician’s orders for oxygen therapy and for routine tubing changes, and staff confirmed the tubing should have been labeled and dated after changes. Resident 29 was observed receiving continuous oxygen at 2 LPM via nasal cannula, but the medical record review failed to show a physician’s order for that continuous oxygen therapy. Staff verified the resident was receiving oxygen and confirmed that an order should have been present. Resident 74 had a physician’s order for continuous oxygen at 2 LPM via nasal cannula, but surveyors observed the resident receiving oxygen at 3 LPM. The resident stated the nurses adjusted the oxygen rate, and a nurse confirmed the resident was receiving 3 LPM throughout the day shift despite the order for 2 LPM. Resident 92 was observed receiving oxygen via nasal cannula with the tubing not on the nares, unlabeled, undated, and without oxygen signage posted on the room door. Resident 137 was observed receiving continuous oxygen at 2 LPM via nasal cannula and using a CPAP machine at night, but there was no oxygen sign outside the room. The CPAP water chamber cleaning instructions in the manufacturer’s guide called for daily cleaning, yet the physician’s orders did not show how often the chamber should be cleaned, and the resident’s care plan did not include the CPAP machine or continuous oxygen therapy. Staff and the DON verified the missing signage, the lack of a care plan for CPAP and oxygen therapy, and the absence of orders addressing CPAP water chamber cleaning.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Incorrect Oxygen Flow Rates and Delayed Tubing Changes
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Two residents receiving O2 had flow rates set above their physician orders, and one resident’s tubing remained dated beyond the weekly change interval. Staff, including an RN and the DON, confirmed the mismatched flow rates and noted that tubing should be changed weekly by the RT. The facility policy required O2 to be administered per physician order and tubing/cannula to be changed weekly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Concentrator Left in Room After Order Discontinued
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Oxygen Concentrator Left in Room After Order Discontinued: A resident with morbid obesity with alveolar hypoventilation and diabetes was observed with an oxygen concentrator at bedside even though there was no current oxygen order. The resident stated he used oxygen only when needed, and the CNO confirmed the oxygen had been discontinued earlier and the concentrator should not have remained in the room.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Order for CPAP Use and Maintenance
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Missing Order for CPAP Use and Maintenance: A resident with sleep apnea and stroke had a CPAP in the room, but there was no physician order or care plan documentation for its use or maintenance. Staff knew the resident had the CPAP, yet CNAs and an RN reported they did not routinely clean the mask or tubing, and the LPN Resident Care Manager confirmed no order existed for the CPAP or its care.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Maintained or Applied as Ordered
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with a continuous O2 order was found without oxygen in place, and an LPN later said the order had been misread. Two other residents with COPD had respiratory equipment that was dirty or outdated, including a gray concentrator filter, tubing and water bottles dated weeks earlier, and tubing lying on the floor with visible dust. Staff gave inconsistent accounts of who was responsible for changing or cleaning the equipment, and the DON acknowledged the dirty equipment and old mask placed residents at risk for respiratory infection or pneumonia.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Not Properly Delivered via Nasal Cannula
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with acute respiratory failure with hypoxia and severe cognitive impairment was ordered oxygen via NC at 2 to 5 LPM to keep O2 saturation at or above 95%. During observation, the resident was lying in bed with the NC on the side of the face instead of in the nose while the concentrator was set at 2.5 LPM. RN and DON stated the NC should be in the nose to deliver oxygen as ordered.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Incorrect Oxygen Flow Settings
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD and chronic respiratory failure did not receive oxygen at the ordered flow rate. Staff found the oxygen concentrator set below the ordered amount on multiple observations, and the resident stated the oxygen was supposed to be set higher. The DON, LVNs, and ADON acknowledged the setting was incorrect and that the ordered flow should have been followed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across California

Get a heads-up on the newest immediate-jeopardy (J–L) citations in California — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.