Respiratory Care Deficiencies With Ventilator Settings, Oxygen Monitoring, and Tubing Dating
Summary
Safe and appropriate respiratory care was not provided for multiple residents who were receiving oxygen therapy or noninvasive ventilation. The facility’s policy required noninvasive ventilation to be provided according to practitioner orders and current standards of practice, with equipment replaced routinely and face mask and tubing changed every 3 months. The oxygen administration policy required oxygen to be administered according to orders and care plans, with monitoring of oxygen saturation levels as ordered and oxygen tubing and mask/cannula changed weekly and as needed, with equipment dated when changed. For one resident who was cognitively intact and receiving a Trilogy noninvasive ventilator at bedtime with heated humidity and 4 liters of oxygen bled in, the order summary did not include settings for the Trilogy device or orders for replacement of the face mask and tubing. The DON confirmed the settings were not obtained from the practitioner and the facility had no record of when the face mask and tubing were last changed. For another resident who was cognitively intact and had an order for oxygen titrated from 2 to 4 liters to keep oxygen saturations greater than 90% continuously via nasal cannula, the record showed only daily oxygen saturation monitoring. The DON confirmed the staff would need to check oxygen saturations more than once a day and at least shiftly to ensure the resident was receiving the correct flow to maintain the ordered saturation level. Two additional residents had oxygen tubing that was not dated. One resident with COPD, obstructive sleep apnea, respiratory failure with hypoxia, and dependence on supplemental oxygen had orders for oxygen at 2 liters per minute via nasal cannula to keep oxygen saturation above 88% and oxygen at 1 liter per minute at night; observations showed oxygen in place, but the tubing had no date marking when it was changed. Another resident with severe cognitive impairment and total dependence for all ADLs had an order for oxygen via nasal cannula at 4 liters per minute, and observations showed the tubing was not dated. The DON confirmed the oxygen tubing was not dated and should be dated when changed, changed weekly, and documented on the TAR.
Penalty
Resources
Below are regulatory guidelines relevant to this citation:
Trusted data from CMS and state health departments
Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release October 8, 2026) and official state health department websites — never guesswork.
In your survey window? See what surveyors are citing.
The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.