F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
D

Bed Rail Assessments and Consents Not Completed

Delaware Veterans HomeMilford, Delaware Survey Completed on 03-12-2026

Summary

The facility failed to assess the entrapment risk of bed rails used for mobility assistance and failed to obtain consents for three residents reviewed for accident hazards: R19, R24, and R37. The report states that the facility also did not have a policy for bed rails. During review of R19, who had diagnoses of Alzheimer's disease with late onset and severe dementia with behavioral disturbances and a BIMS score of 0, the resident was observed in bed with 1/4 side rails up on both sides, but there was no evidence in the EMR of any bed rail assessment, discussion of risks and benefits, or signed consent from the resident's representative. RN1 later observed that R19 could not grab or use the side rails for mobility or during care and stated the resident had never been able to use them. For R24, who was admitted with a left femur fracture and had BIMS scores of 13 and later 15, one bed rail assessment stated the resident did not need bed rails and was signed by the resident, but a later assessment was blank. Despite the earlier assessment indicating bed rails were not indicated, the resident was observed with 1/4 side rails up on both sides of the bed and said he used them for positioning, but could not remember being informed of the risks and benefits. For R37, who had hemiplegia and hemiparesis following a stroke and a BIMS score of 15, the bed rail assessment was blank, yet the resident was observed with bilateral 1/4 side rails up and stated he used them for positioning and mobility, but could not remember if anyone had reviewed the risks and benefits or when the rails were placed on the bed. The Administrator stated she thought 1/4 side rails did not require consent or explanation of risks and benefits, and confirmed there were no signed consents or evidence that R19, R24, or R37 or their representatives were provided the risks and benefits.

Penalty

Inspection fine: $86,920
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0700 citations
Failure to Assess and Obtain Consent Before Bed Rail Use
D
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

Failure to assess and obtain consent before bed rail use for two residents. Two cognitively intact residents with diagnoses including AFib, HTN, arthritis, CAD, and HF were observed with bed rails on their beds even though their MDSs indicated no bedrail use. Both records lacked a bedrail assessment, consent, and education, and the DON confirmed no assessment had been completed before the rails were installed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess Bed Rail Entrapment Risk and Mattress Settings
J
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A resident with severe cognitive impairment and total dependence for care fell from bed after an air mattress was set too high for the resident’s weight, causing the mattress to overinflate and push the resident into a siderail. The resident’s arm became trapped in the rail and the resident sustained neck and sternum injuries. The record also showed no siderail entrapment assessment for the resident, and multiple other residents with air mattresses and siderails also lacked documented side rail assessments.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Reassess Bed Siderail Safety Led to Resident Injury
G
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A facility failed to complete ongoing reassessments of bed siderail safety for a resident with severe cognitive impairment and total dependence for bed mobility. The resident continued using the siderails for months without the required review, and later developed bruising and a traumatic posterior shoulder dislocation after the arm was reportedly caught in the rail. An internal investigation and later assessment found the siderails were unsafe and no longer indicated.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess and Obtain Consent for Bed Rail Use
D
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

Failure to Assess and Obtain Consent for Bed Rail Use: Two residents had one-quarter bed rails in place with physician orders for positioning and mobility, but the clinical record lacked documented bed rail safety assessments and signed informed consent. One resident had moderately impaired cognition and the other had severely impaired cognition; both were observed with the rails up and locked in place. The ADM and DON acknowledged the assessments and consents were missed, and staff stated the required documentation should have been completed before bed rail use.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Document Alternatives and Monitor Bed Rail Use
D
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A resident with acute and chronic respiratory failure with hypoxia, Type 2 DM with hyperglycemia, generalized muscle weakness, and impaired cognition was placed in four locked bed rails for fall risk. The record showed only one alternative intervention, one-to-one activities and supervision, with no documented evidence explaining why it was ineffective, and the DON stated there was no documented active order for a low bed or other alternative. The resident was observed asleep with all four bed rails up and locked, and the care plan and active orders did not show monitoring for safe bed rail use.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess Bed Rail Entrapment Risk
J
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

Failure to Assess Bed Rail Entrapment Risk: A resident with severe cognitive impairment and multiple medical conditions was placed in a bed with assist rails without documented assessment for appropriateness, documented alternatives, or informed consent. Staff later found the resident with his head and neck trapped between the mattress and the assist bar, unresponsive and not breathing, and CPR was started before EMS transported him to the hospital. The resident was admitted with cardiac arrest and diffuse hypoxic injury and later expired.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Delaware

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Delaware — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.