Dusty Oxygen Concentrator Filters
Summary
The facility failed to ensure necessary respiratory care and services were provided in accordance with professional standards of practice for 2 of 5 residents reviewed for respiratory care. Resident 27 was observed on multiple occasions wearing oxygen at 2 LPM via nasal cannula, and the filter on the oxygen condenser was observed to be dusty each time. Resident 27’s record showed diagnoses including hemiplegia and hemiparesis following cerebral infarction affecting the right dominant side and hypoxia, and the most recent MDS indicated moderately impaired cognition and oxygen therapy use. The physician’s orders included continuous humidified supplemental oxygen at 2-4 LPM to maintain O2 saturation above 90%, monitoring O2 saturation every shift, and wiping down the portable oxygen tank and room concentrator with tubing and humidifier changes every Sunday, but the record lacked a physician’s order related to oxygen concentrator filters. The care plan for hypoxia also lacked an intervention related to oxygen concentrator filters. Resident 9 was also observed wearing oxygen at 2 LPM via nasal cannula, and the filter on the back of the oxygen condenser was observed to be dusty on more than one occasion. Resident 9’s record showed chronic respiratory failure with hypoxia, and the most recent MDS indicated the resident was cognitively intact and receiving oxygen therapy. The physician’s orders included changing the O2 tubing and prefilled humidifier every Sunday, but the clinical record lacked a physician’s order related to oxygen concentrator filters, and the impaired gas exchange care plan lacked an intervention related to oxygen concentrator filters. During interviews, an LPN stated maintenance cleaned the oxygen filters on the oxygen condensers, while a maintenance assistant and central supply indicated the oxygen provider cleaned the filters twice a year per the provider’s policy. The administrator reviewed the contract with the oxygen provider, which stated supplies and services would meet industry standards, but specific information related to cleaning oxygen condenser filters was lacking, and a policy for cleaning oxygen condenser filters was requested and not received.
Penalty
Resources
Below are regulatory guidelines relevant to this citation:
Trusted data from CMS and state health departments
Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release July 29, 2026) and official state health department websites — never guesswork.
In your survey window? See what surveyors are citing.
The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.