F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
D

Improper Storage of Nasal Cannula

Carlyle Senior Care Of AikenAiken, South Carolina Survey Completed on 09-12-2025

Summary

The facility failed to ensure a resident's nasal cannula was stored properly when not in use for one of two residents reviewed for oxygen therapy, Resident 65. The facility's Oxygen Safety policy dated 04/2025 stated it was the policy of the facility to provide a safe environment for residents, staff, and the public and addressed the use and storage of oxygen and oxygen equipment. Resident 65 was readmitted to the facility with diagnoses including quadriplegia, had a BIMS score of 15 out of 15 on the quarterly MDS dated 08/06/25, and was coded as receiving oxygen therapy. The physician order dated 02/27/25 ordered oxygen at 2 liters per minute via nasal cannula as needed for shortness of breath. During observation on 09/11/25 at 2:35 PM, Resident 65's oxygen nasal cannula was wrapped around the top of the resident's oxygen concentrator, and on 09/12/25 at 12:00 PM it was observed lying at the foot of the resident's bed uncovered. During interview, LPN3 verified the cannula was left out uncovered and stated it should be stored in a bag. Later, LPN4 observed the cannula lying on the floor uncovered and stated it should not be on the floor and should be stored in a bag. The DON also stated that a resident's nasal cannula should be stored in a bag when not in use.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Oxygen equipment not maintained per order
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident who required oxygen therapy for SOB had an oxygen bubbler that was overdue for replacement, despite a TAR directing weekly changes of oxygen supplies including the bubbler and tubing. Staff confirmed the bubbler should have been changed weekly, and the DON and staff educator stated nurses were expected to follow the provider order as written and that all disposable oxygen pieces were to be changed weekly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Bi-pap Therapy Not Ordered or Verified at Admission
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD, acute respiratory failure with hypercapnia, and oxygen dependence was admitted with prior records showing nightly Bi-pap use and specific settings, but the facility did not have a Bi-pap order or settings in the chart and the baseline care plan did not address oxygen or Bi-pap. The resident reported he went without Bi-pap for three nights because a connector piece was missing, while the hospice RN, admitting LPN, ADON, and DON each stated they were unaware of the machine or did not verify the equipment and orders at admission.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Care and Tracheostomy Care Not Provided as Ordered
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory care was not provided as ordered for one resident receiving O2, as the NC tubing was not dated. Tracheostomy care was also not provided as ordered for another resident: heavy thick mucus was visible under the trach, the ATM mask was soiled with mucus, the ATM and nebulizer tubing were not dated or stored in a bag, and the suction catheter tubing end was not stored in a bag. An RN confirmed the findings.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Change Oxygen Equipment per Physician Orders
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

The facility failed to follow physician orders for oxygen equipment for two residents with COPD-related diagnoses. Observations showed each resident’s O2 tubing and water bottle remained dated and unchanged beyond the ordered weekly schedule, and the DON confirmed the equipment should have been changed per the orders.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Ordered Continuous Oxygen
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Failure to maintain ordered continuous O2: A resident with stroke and COPD was observed without his NC while a CNA waited for a portable O2 unit to be filled. The CNA then applied the NC but did not turn on the liquid portable O2, and later wheeled the resident to the nurse’s station to have the nurse set the unit to the ordered liter flow. The resident’s physician had ordered continuous O2 2L to 4L NC.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage of CPAP Mask and Oxygen Tubing
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Improper Storage of CPAP Mask and Oxygen Tubing: A resident with COPD, DM, dialysis dependence, and continuous O2 orders had his CPAP/NIV mask observed lying on the bedside table and his O2 tubing wrapped under the concentrator handle instead of being stored in a sanitary container or bag. Staff interviews confirmed the equipment should be stored in a bag when not in use, but it was not being kept that way during observation.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across South Carolina

Get a heads-up on the newest immediate-jeopardy (J–L) citations in South Carolina — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release October 8, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.