F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
D

Failure to Maintain Respiratory Equipment Cleanliness

Terrace View Care CenterFullerton, California Survey Completed on 02-27-2025

Summary

The facility failed to provide safe respiratory care services for two residents, Resident 8 and Resident 294, by not adhering to the manufacturer's guidelines and the facility's policies and procedures for cleaning and maintaining respiratory equipment. For Resident 8, the facility did not ensure that the non-invasive ventilator machine was cleaned according to the manufacturer's guidelines, nor did they ensure that the headgear and tubing were cleaned as per the facility's policies. Observations revealed that the nasal cannula tubing and BiPAP oxygen tubing were touching the ground, and there was no documentation of the cleaning of the BiPAP machine or the filter being checked. Resident 8, who was admitted with a diagnosis of COPD and required a non-invasive mechanical ventilator, had a care plan that failed to include interventions for the care and maintenance of the BiPAP device. Despite physician's orders for cleaning the headgear and tubing, there was no evidence that these were followed consistently. Interviews with staff, including an LVN and the DON, confirmed the lack of documentation and adherence to cleaning protocols, with the DON acknowledging the findings. For Resident 294, the facility did not ensure that the CPAP machine was cleaned according to the manufacturer's guidelines, and the mask and tubing were not stored properly when not in use. Observations showed the CPAP mask and tubing were placed on top of a drawer with the tubing touching the floor. The resident, who had a sleep disorder, was independent in using the CPAP machine, but there was no physician's order for cleaning the device weekly as directed by the user guide. Interviews with staff, including a CNA and an LVN, revealed that the licensed nurses were responsible for cleaning the CPAP machine, but there was no documentation of this being done, and the DON acknowledged the lack of compliance with the manufacturer's cleaning instructions.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Incorrect Oxygen Flow Rates and Delayed Tubing Changes
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Two residents receiving O2 had flow rates set above their physician orders, and one resident’s tubing remained dated beyond the weekly change interval. Staff, including an RN and the DON, confirmed the mismatched flow rates and noted that tubing should be changed weekly by the RT. The facility policy required O2 to be administered per physician order and tubing/cannula to be changed weekly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Concentrator Left in Room After Order Discontinued
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Oxygen Concentrator Left in Room After Order Discontinued: A resident with morbid obesity with alveolar hypoventilation and diabetes was observed with an oxygen concentrator at bedside even though there was no current oxygen order. The resident stated he used oxygen only when needed, and the CNO confirmed the oxygen had been discontinued earlier and the concentrator should not have remained in the room.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Order for CPAP Use and Maintenance
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Missing Order for CPAP Use and Maintenance: A resident with sleep apnea and stroke had a CPAP in the room, but there was no physician order or care plan documentation for its use or maintenance. Staff knew the resident had the CPAP, yet CNAs and an RN reported they did not routinely clean the mask or tubing, and the LPN Resident Care Manager confirmed no order existed for the CPAP or its care.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Maintained or Applied as Ordered
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with a continuous O2 order was found without oxygen in place, and an LPN later said the order had been misread. Two other residents with COPD had respiratory equipment that was dirty or outdated, including a gray concentrator filter, tubing and water bottles dated weeks earlier, and tubing lying on the floor with visible dust. Staff gave inconsistent accounts of who was responsible for changing or cleaning the equipment, and the DON acknowledged the dirty equipment and old mask placed residents at risk for respiratory infection or pneumonia.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Not Properly Delivered via Nasal Cannula
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with acute respiratory failure with hypoxia and severe cognitive impairment was ordered oxygen via NC at 2 to 5 LPM to keep O2 saturation at or above 95%. During observation, the resident was lying in bed with the NC on the side of the face instead of in the nose while the concentrator was set at 2.5 LPM. RN and DON stated the NC should be in the nose to deliver oxygen as ordered.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Incorrect Oxygen Flow Settings
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD and chronic respiratory failure did not receive oxygen at the ordered flow rate. Staff found the oxygen concentrator set below the ordered amount on multiple observations, and the resident stated the oxygen was supposed to be set higher. The DON, LVNs, and ADON acknowledged the setting was incorrect and that the ordered flow should have been followed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across California

Get a heads-up on the newest immediate-jeopardy (J–L) citations in California — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.