F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
E

Failure to Ensure Proper Installation and Testing of Bed Rails

Complete Care At Maple Grove LlcMadison, Wisconsin Survey Completed on 11-11-2024

Summary

The facility failed to ensure the correct installation, use, and maintenance of bed rails for four residents, identified as R3, R6, R7, and R8. The deficiency was identified through observation, interviews, and record reviews, revealing that bed rails were installed without conducting a Bed System Measurement Device Test to ensure proper installation and reduce the risk of entrapment. The facility's policy requires that physical devices, including bed rails, be reviewed for safety and used according to the manufacturer's recommendations, with regular inspections to prevent safety hazards. For resident R3, the surveyor observed a gap between the bed rail and the mattress, indicating improper installation. The Maintenance Supervisor (MS C) admitted that a newly hired employee, who had not been trained on the measurement device, was installing bed rails. MS C also acknowledged that several installations had not been tested for entrapment points. The facility provided a Bed System Measurement Device Test for R3 dated after the surveyor's request, with no prior documentation of testing. Similarly, residents R6, R7, and R8 had bed rails installed without prior testing documentation. The facility's records indicated order dates for the bed rails, but no evidence of testing before the surveyor's intervention. The Nursing Home Administrator (NHA A) confirmed that testing should occur upon installation, but the facility lacked documentation of actual installation dates, relying instead on order dates. This oversight in testing and documentation led to the deficiency identified by the surveyors.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0700 citations in Ohio
Failure to Assess Residents Prior to Bed Rail Use
E
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

Surveyors determined that the facility failed to assess multiple residents for the appropriateness of bed rail use before installing bed rails on their beds. Observation with the DON revealed numerous residents with bed rails in place, and the DON confirmed that no prior safety risk assessments or evaluations of less restrictive alternatives had been completed, despite a written policy requiring such assessments and documentation before bed rails are used.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Bed Rail Installation Resulting in Resident Injury
G
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A resident with significant mobility and medical needs was injured when a bed rail detached during in-bed care, causing a fall and a displaced upper arm fracture. Staff interviews and documentation revealed the bed rails had been installed incorrectly on the bed frame, were previously reported as loose, and were not compatible with the bed's crossbar. The facility lacked the correct user manual for the bed rails, and staff had previously adjusted the rails improperly, leading to the incident.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess and Mitigate Bed Rail Entrapment Risks Results in Resident Death and Immediate Jeopardy
J
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A facility failed to properly assess and document the risks of bed rail entrapment for residents using alternating pressure mattresses, leading to a resident's death by asphyxia after becoming wedged between the mattress and bed rail. The facility did not measure mattress gaps under compression, did not document medical need or alternatives to bed rails, and did not attempt alternative interventions before installing side rails for multiple residents, placing several at risk for harm.

Fine: $26,685
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess Bed Rail Risks for a Resident
D
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A facility failed to assess a resident for entrapment risks before installing bed rails, despite the resident's medical conditions such as hemiplegia and seizures. The resident's care plan included bed rails due to fall risk, but no assessment was documented. The facility's policy requires such assessments, which were not conducted.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess and Document Bed Rail Use
D
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A facility failed to assess and document the need for bed rails for a resident receiving hospice care. Despite the use of bed rails being noted in a consent form, there were no physician orders or assessments, and the form lacked necessary signatures. Observations confirmed the use of bed rails, but they were not coded in the MDS assessments. The facility's policy required assessments and evaluations that were not completed, leading to the deficiency.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess and Obtain Consent for Bed Rail Use
E
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

The facility failed to assess and obtain consents or orders for bed rail use for six residents, despite their need for extensive ADL assistance. The facility's policy requires assessments and informed consent, but these were not documented. The DON confirmed the absence of necessary records, indicating a systemic compliance failure.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

65.1% of Ohio facilities received at least one citation during their inspection in the last 12 months.Will yours be survey-ready?

Surveyors issued 55 serious citations across Ohio in the last 12 months. See exactly what they're citing.

Get ready for your next survey

See what surveyors are citing in Ohio and spot your risk areas before they do.

Monthly Citation Reports

Have you been cited for this tag?

Save hours drafting a compliant Plan of Correction — AI built on real approved POCs.

Plan of Correction Writer

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release May 27, 2026) and official state health department websites — never guesswork.

Trusted by long-term care providers and associations.

Allegria Senior Living logo
FHCA logo
WeCare Centers logo
Care Rehab logo
An unhandled error has occurred. Reload 🗙