F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
D

Deficiencies in Respiratory Care and Infection Control

Encore At Boca Raton Rehabilitation And Nursing CeBoca Raton, Florida Survey Completed on 12-05-2024

Summary

The facility failed to ensure that a resident receiving oxygen had a physician's order, affecting one of the four sampled residents reviewed for respiratory care. Specifically, Resident #111 was observed wearing oxygen without any evidence of a physician's order for oxygen administration. This oversight was noted during an observation on December 2, 2024, when the resident was seen sitting up in bed with an oxygen concentrator set at 2 liters. The resident had been admitted with diagnoses including acute and chronic respiratory failure with hypoxia and chronic obstructive pulmonary disease. Additionally, the facility did not maintain respiratory supplies in a manner that prevents infection for four sampled residents. Resident #6's nebulizer mask was observed on top of a personal cart and later on a nightstand, not stored in a plastic bag as required. Resident #17's oxygen tubing had not been changed since November 11, 2024, despite a physician's order for weekly changes. Resident #8's oxygen cannula and tubing were observed with a date tag of November 18, 2024, and were found on the floor, not stored in any plastic containment. Interviews with the residents and the Director of Nursing (DON) revealed inconsistencies in the facility's adherence to infection control protocols. The DON confirmed that residents need a physician's order for oxygen and that oxygen tubing should be changed weekly. The DON also stated that nebulizer masks should be stored in plastic bags when not in use. These observations and interviews highlight the facility's failure to follow its own infection prevention policies related to respiratory care equipment.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Incorrect Oxygen Flow Rates and Delayed Tubing Changes
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Two residents receiving O2 had flow rates set above their physician orders, and one resident’s tubing remained dated beyond the weekly change interval. Staff, including an RN and the DON, confirmed the mismatched flow rates and noted that tubing should be changed weekly by the RT. The facility policy required O2 to be administered per physician order and tubing/cannula to be changed weekly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Concentrator Left in Room After Order Discontinued
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Oxygen Concentrator Left in Room After Order Discontinued: A resident with morbid obesity with alveolar hypoventilation and diabetes was observed with an oxygen concentrator at bedside even though there was no current oxygen order. The resident stated he used oxygen only when needed, and the CNO confirmed the oxygen had been discontinued earlier and the concentrator should not have remained in the room.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Order for CPAP Use and Maintenance
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Missing Order for CPAP Use and Maintenance: A resident with sleep apnea and stroke had a CPAP in the room, but there was no physician order or care plan documentation for its use or maintenance. Staff knew the resident had the CPAP, yet CNAs and an RN reported they did not routinely clean the mask or tubing, and the LPN Resident Care Manager confirmed no order existed for the CPAP or its care.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Maintained or Applied as Ordered
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with a continuous O2 order was found without oxygen in place, and an LPN later said the order had been misread. Two other residents with COPD had respiratory equipment that was dirty or outdated, including a gray concentrator filter, tubing and water bottles dated weeks earlier, and tubing lying on the floor with visible dust. Staff gave inconsistent accounts of who was responsible for changing or cleaning the equipment, and the DON acknowledged the dirty equipment and old mask placed residents at risk for respiratory infection or pneumonia.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Not Properly Delivered via Nasal Cannula
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with acute respiratory failure with hypoxia and severe cognitive impairment was ordered oxygen via NC at 2 to 5 LPM to keep O2 saturation at or above 95%. During observation, the resident was lying in bed with the NC on the side of the face instead of in the nose while the concentrator was set at 2.5 LPM. RN and DON stated the NC should be in the nose to deliver oxygen as ordered.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Incorrect Oxygen Flow Settings
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD and chronic respiratory failure did not receive oxygen at the ordered flow rate. Staff found the oxygen concentrator set below the ordered amount on multiple observations, and the resident stated the oxygen was supposed to be set higher. The DON, LVNs, and ADON acknowledged the setting was incorrect and that the ordered flow should have been followed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Florida

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Florida — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release August 26, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.