F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
E

Deficiencies in Bed Rail Policy Adherence

The Villas At Saratoga Skilled Nsg & Assisted LvgSaratoga, California Survey Completed on 12-20-2024

Summary

The facility failed to adhere to its bed rail policy for 12 out of 18 sampled residents, leading to deficiencies in the management and documentation of bed rail use. Specifically, several residents were using grab bars without the necessary physician orders or care plans. For instance, Resident 2 had grab bars installed without a physician order, despite a bed rail use evaluation recommending their use for mobility and positioning assistance. Similarly, Residents 43 and 67 were observed with grab bars in use, but there were no corresponding physician orders documented. Additionally, the facility did not have care plans in place for the use of grab bars for several residents, including Residents 8, 23, 33, and 82, even though physician orders were present. This lack of care planning was confirmed during interviews with the Director of Nursing (DON), who acknowledged the absence of care plans and stated that they should have been in place. Furthermore, Resident 70 was using grab bars without both a physician order and a care plan until these were initiated during the survey period. The report also highlights that Residents 4, 16, 20, and 51 had grab bars in use without the necessary care plans, and in some cases, without physician orders. The DON and the Minimum Data Set Director confirmed these omissions during interviews. The facility's policy on the use of side rails/bed rails emphasizes the need for these to be addressed in the resident's care plan, which was not adhered to in these cases, potentially compromising resident safety.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0700 citations in Ohio
Failure to Assess Residents Prior to Bed Rail Use
E
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

Surveyors determined that the facility failed to assess multiple residents for the appropriateness of bed rail use before installing bed rails on their beds. Observation with the DON revealed numerous residents with bed rails in place, and the DON confirmed that no prior safety risk assessments or evaluations of less restrictive alternatives had been completed, despite a written policy requiring such assessments and documentation before bed rails are used.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Bed Rail Installation Resulting in Resident Injury
G
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A resident with significant mobility and medical needs was injured when a bed rail detached during in-bed care, causing a fall and a displaced upper arm fracture. Staff interviews and documentation revealed the bed rails had been installed incorrectly on the bed frame, were previously reported as loose, and were not compatible with the bed's crossbar. The facility lacked the correct user manual for the bed rails, and staff had previously adjusted the rails improperly, leading to the incident.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess and Mitigate Bed Rail Entrapment Risks Results in Resident Death and Immediate Jeopardy
J
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A facility failed to properly assess and document the risks of bed rail entrapment for residents using alternating pressure mattresses, leading to a resident's death by asphyxia after becoming wedged between the mattress and bed rail. The facility did not measure mattress gaps under compression, did not document medical need or alternatives to bed rails, and did not attempt alternative interventions before installing side rails for multiple residents, placing several at risk for harm.

Fine: $26,685
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess Bed Rail Risks for a Resident
D
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A facility failed to assess a resident for entrapment risks before installing bed rails, despite the resident's medical conditions such as hemiplegia and seizures. The resident's care plan included bed rails due to fall risk, but no assessment was documented. The facility's policy requires such assessments, which were not conducted.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess and Document Bed Rail Use
D
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

A facility failed to assess and document the need for bed rails for a resident receiving hospice care. Despite the use of bed rails being noted in a consent form, there were no physician orders or assessments, and the form lacked necessary signatures. Observations confirmed the use of bed rails, but they were not coded in the MDS assessments. The facility's policy required assessments and evaluations that were not completed, leading to the deficiency.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess and Obtain Consent for Bed Rail Use
E
F0700 F700: Try different approaches before using a bed rail. If a bed rail is needed, the facility must (1) assess a resident for safety risk; (2) review these risks and benefits with the resident/representative; (3) get informed consent; and (4) Correctly install and maintain the bed rail.
Short Summary

The facility failed to assess and obtain consents or orders for bed rail use for six residents, despite their need for extensive ADL assistance. The facility's policy requires assessments and informed consent, but these were not documented. The DON confirmed the absence of necessary records, indicating a systemic compliance failure.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

65.1% of Ohio facilities received at least one citation during their inspection in the last 12 months.Will yours be survey-ready?

Surveyors issued 55 serious citations across Ohio in the last 12 months. See exactly what they're citing.

Get ready for your next survey

See what surveyors are citing in Ohio and spot your risk areas before they do.

Monthly Citation Reports

Have you been cited for this tag?

Save hours drafting a compliant Plan of Correction — AI built on real approved POCs.

Plan of Correction Writer