F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
D

Failure to Conduct Routine Bed Rail Inspections

Lakeview Post AcuteFlorissant, Missouri Survey Completed on 10-21-2024

Summary

The facility failed to ensure routine inspections of bed/side rails were completed as part of a regular maintenance program, leading to potential entrapment risks for three residents. The FDA guidance suggests evaluating the dimensional limits of gaps in hospital beds to mitigate entrapment risks, especially for vulnerable populations such as the elderly. The facility's policy requires maintenance staff to routinely inspect beds and related equipment, with results reported to the Quality Assurance and Performance Improvement (QAPI) committee. However, the facility did not adhere to these guidelines, as evidenced by the lack of documentation and assessments for the residents involved. Resident #47, who had moderate cognitive impairment and no documented use of side rails, was observed with bilateral side rails raised on multiple occasions. Despite the resident's use of side rails for positioning, there was no physician order or care plan addressing this use. Similarly, Resident #27, with severe cognitive impairment and dependence on assistance for mobility, had a care plan that included the use of side rails for bed mobility. However, the evaluation of entrapment risk did not include measurements, and observations showed the resident with top quarter rails up. Resident #344, who was alert with limited speech and had multiple diagnoses including a traumatic brain injury, was observed with bilateral quarter side rails raised. There was no documentation of a Nursing Bed Rail Observation/Assessment or physician order for side rails. Interviews with staff revealed that the maintenance department was responsible for safety checks, but there was no documentation of siderail measurements being completed, despite the Maintenance Director's awareness of entrapment risks and possession of the necessary tools. The Administrator expected siderails to be measured and checked for entrapment quarterly, but this was not documented.

Penalty

9 days payment denial
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0909 citations
Missing Bed Rail Inspection Documentation
E
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Missing Bed Rail Inspection Documentation: The facility failed to complete and/or maintain routine bed rail inspection records for all residents with side rails. A resident with bilateral half rails used for repositioning and safety due to frequent falls out of bed was observed with upper quarter-length rails, and maintenance staff from a sister facility later checked the rails without prior facility documentation. An undated bed entrapment inspection form showed 10 beds passed, but it did not identify who completed the assessments.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Unsafe Bed Frame, Mattress, and Bed Rail Conditions
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Unsafe Bed Frame, Mattress, and Bed Rail Conditions: Two residents had unsafe bed setups. One resident with CHF, obesity, chronic respiratory failure, and dependence for bed mobility had an undersized mattress that exposed metal bed frame and left gaps between the mattress and quarter side rail; when turned, the resident’s foot contacted the sharp metal frame. Another resident with dementia, schizophrenia, prior CVA/TIA, and dependence for bed mobility had a loose left quarter rail and no right rail, despite orders for bilateral upper quarter rails.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Nonfunctional Bed Frames and Electric Bed Not in Good Repair
E
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Nonfunctional Bed Frames and Electric Bed Not in Good Repair: The facility failed to ensure that several residents' bed frames had working manual lift mechanisms and that one resident's electric bed was in good repair. During observation, an electric bed control would not function even after the cord was found and reconnected, and multiple manual crank bed frames for other residents were observed not to work properly. A CNA stated the crank handles could not lift the beds, and the Maintenance Director acknowledged some crank handles were rusted and had been repaired with oil spray.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Inspect U-Bar as Part of Regular Bed Safety Program
E
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Failure to Inspect U-Bar as Part of Regular Bed Safety Program: A resident with severe cognitive impairment and transfer assistance needs had a u-bar on the bed for transfers, but the facility did not include the u-bar in its regular maintenance inspection program. Surveyors observed the u-bar in place, reviewed records showing the device was ordered and used for transfers, and found no evidence that maintenance had regularly inspected it. The maintenance supervisor stated u-bars were not routinely inspected, and the DON/administrator were unaware of why regular inspections were not being done.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Conduct and Document Routine Bed Inspections
F
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Facility staff failed to conduct and document routine inspections of resident bed frames, mattresses, and bed rails. Surveyors reviewed maintenance logs and found they reflected various repairs but did not show the required monthly bed inspections. The Administrator stated that routine inspections were not being conducted, and the DON provided a policy that included a 12-point monthly bed inspection checklist.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Bed Mattresses Did Not Fit Bed Frames, Creating Entrapment Gaps
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Bed mattresses did not fit the bed frames for three residents, leaving exposed gaps at the foot of the bed and between the mattress and side rail. One resident reported a foot had gotten caught in the gap more than once, another had a 7-inch gap at the head and side of the bed, and review of bed assessments showed the foot end had not been assessed for possible entrapment.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Missouri

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Missouri — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release July 29, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙