F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
D

Improper Oxygen Administration Due to Unupdated Physician Order

Saginaw Senior Care And Rehabilitation Center, LlcSaginaw, Michigan Survey Completed on 10-09-2024

Summary

The facility failed to revise and update a physician's order for oxygen administration for a resident, resulting in the improper flow rate of oxygen being administered. The resident, who is severely cognitively impaired with a BIMS score of 0, was observed on multiple occasions receiving oxygen at a flow rate of 4 liters per minute via a nasal cannula, despite the physician's order specifying a flow rate of 2 liters per minute. This discrepancy was noted during observations on two consecutive days, with the oxygen tubing dated from the day before the first observation. The issue was brought to the attention of the Clinical Care Coordinator, who verified the incorrect flow rate and indicated they would check the resident's oxygen level and notify the physician. The facility's policy on oxygen administration emphasizes the importance of adhering to physician orders and includes procedures for ensuring safe oxygen therapy. However, the failure to follow the physician's order for the correct oxygen flow rate led to the deficiency identified in the report.

Penalty

Inspection fine: $51,282
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Missing Oxygen Order and Improper Nebulizer Storage
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Missing Oxygen Order and Improper Nebulizer Storage: A resident receiving oxygen for sleep apnea had no physician order specifying the oxygen delivery rate, even though staff were setting the rate. In a separate observation, a resident’s nebulizer mask was left unbagged on the nightstand while not in use. An LPN and the DON both stated respiratory equipment should be stored in a bag when not in use, and the facility policy required bagged storage between uses.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory equipment was not maintained and stored per standards
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory care was not provided in accordance with standards for several residents receiving oxygen and nebulizer treatments. Staff observed outdated oxygen tubing and humidification equipment, a nebulizer mouthpiece left out with tubing touching the floor, and oxygen supplies not stored properly. One resident’s oxygen was running at a higher flow than ordered, and staff confirmed the tubing and humidification items should be changed weekly and documented.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage and Dating of Oxygen and CPAP Equipment
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Improper Storage and Dating of Oxygen and CPAP Equipment: Two residents had oxygen equipment and CPAP items observed out of proper storage, including nasal cannula tubing on the floor, undated or empty humidifiers, and a CPAP mask on the floor. One resident had OSA with an order for oxygen at bedtime and as needed, and the other had COPD with orders for nightly CPAP and continuous oxygen. Staff stated the tubing, humidifier, and CPAP mask should be stored and dated per facility practice, and the facility policy required weekly changes and dating of oxygen equipment.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Contaminated oxygen tubing was placed back on a resident
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident receiving continuous O2 via nasal cannula had the cannula found on the floor while the concentrator was running. A CNA placed the cannula back on the resident’s face before an LVN identified it as contaminated and replaced the cannula and tubing. The resident had diagnoses including hypertensive heart disease and atherosclerotic heart disease, and the care plan called for O2 at 2 L/min to maintain O2 sats above 92%.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Stored Sanitarily
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD, a respiratory infection, continuous O2, and CPAP therapy had CPAP and nasal cannula tubing observed lying on the bed and the CPAP mask draped over the machine instead of being stored in a sanitary container. Staff interviews showed inconsistent understanding of how to store the respiratory equipment, and the facility did not provide the requested storage of the equipment when not in use.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Properly Labeled or Stored
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory equipment was not properly changed, labeled, or stored for a resident receiving nebulizer treatment. An undated aerosol mask attached to a nebulizer was observed on the resident’s nightstand, open to air, and remained there on a later observation. An LPN confirmed the mask was open to air and not stored properly, and the DON confirmed the mask and tubing had not been dated or stored properly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Michigan

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Michigan — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release July 29, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙