F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
E

Deficient Call Light System in LTC Facility

Gardens Nursing And Rehab CenterMiami, Florida Survey Completed on 08-30-2024

Summary

The facility failed to ensure that residents had a functioning call light system to request assistance from staff, affecting five residents. The facility's policy requires that call lights be answered promptly and any malfunctions reported to maintenance or relevant staff. However, several residents reported that their call lights were not working, and staff were aware of these issues but did not resolve them. Resident #6, who is cognitively intact and requires assistance with personal care, reported being left soaking wet in bed due to a non-functioning call light. Resident #9, also cognitively intact, stated that her call light had been malfunctioning for over six months, and despite having manual call bells, staff could not hear them. Resident #19, who uses a wheelchair and requires supervision for toilet transfers, reported that his call light had not worked for over a month. Maintenance staff acknowledged the issue but indicated they were not equipped to fix it. Observations confirmed that the call light system in Resident #19's bathroom did not function, as no auditory signal was heard at the nurses' station. Resident #83 demonstrated that their call light did not activate any notification at the nurses' station, and no staff responded to the call light during the observation period. Similarly, Resident #88's call light did not trigger any notification, and no staff responded when it was activated. Interviews with staff revealed that the call light system had been malfunctioning for an extended period, with one CNA stating that the issue had persisted for the entire six months of her employment. The facility administrator acknowledged that repairs were underway but indicated that the issue could not be resolved immediately due to the need for specific parts. This ongoing deficiency in the call light system compromised the residents' ability to communicate their needs effectively to the staff.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0919 citations
Bathroom Emergency Call Light Pull Cords Not Accessible
E
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Bathroom emergency call light pull cords in several resident rooms were observed hanging above the handrail and not readily accessible from the floor. The Maintenance Director stated they were unaware of any regulation related to bathroom pull cords, and the Administrator stated the expectation was that resident bathroom pull cords should be accessible if a resident was on the floor.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Nonfunctioning Resident Call Lights
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Nonfunctioning Resident Call Lights: A resident reported waiting a long time for staff to answer her call light and said it had been disconnected from the wall extender, while another resident said her call light had been unplugged and she had to get help by having a friend use his call light. Staff and residents also reported repeated call light problems for another resident, and the DON stated call lights were expected to always be in working order; the facility did not have a call light policy.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Nonfunctioning Resident Call Buttons
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Nonfunctioning Resident Call Buttons: Two residents had call buttons that did not work, and testing showed no light or alarm at the nurse’s station. An LVN later gave each resident a handbell as a temporary replacement, but one resident reported ringing it overnight without response and woke up with a soaked brief. Staff interviews showed the handbells were not clearly communicated to all aides, and the MAINTD said the call system problem was related to cords wrapped around bed frames causing a short.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Call Light Not Accessible in Shower Area
E
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

A resident with hemiplegia/hemiparesis, HF, gait impairment, and muscle weakness required max assist with showering and toilet hygiene, and the care plan included keeping the call light within reach. During observation, the resident was seated in a shower chair while an NA left the room, and the bathroom call light box was located across the room next to the toilet, making it unreachable from the shower. Staff and the DON stated residents were not to be left unattended in the shower, but the new call light setup left no accessible call light or cord in the shower area despite the facility policy requiring call lights at the bedside, toilet, and shower/bathing facility.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Nonfunctioning Call Light System on Cove Unit
E
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

A resident call light system on the Cove unit was not working, and multiple residents reported they could not summon help when needed. Residents said their call lights did not illuminate or sound, some had to yell, wheel into the hallway, or wait for staff rounds, and not all residents were given bells. Staff knew parts of the system were down, but communication and response were inconsistent, and maintenance was not notified immediately.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Call Light Not Available Within Residents’ Reach
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Call Light Not Within Reach for Two Residents: A facility failed to ensure a working call system was available in resident rooms and bathing areas. Two cognitively intact residents who needed assistance with ADLs reported they had no call light or bell to summon staff, and observations confirmed no call light or bell was present in their rooms. The facility policy required call lights to be within the resident’s reach at all times.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Florida

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Florida — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release July 29, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙