F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
D

Deficiencies in Oxygen Administration for Residents

Pioneer Health Care CenterRocky Ford, Colorado Survey Completed on 08-29-2024

Summary

The facility failed to provide proper respiratory care for three residents who required supplemental oxygen. Resident #3, who had a diagnosis of COPD, was observed with an empty portable oxygen concentrator, despite a physician's order for continuous oxygen at 2 LPM. The resident's oxygen saturation level was at 90%, and she reported feeling 'loopy,' indicating potential hypoxia. The staff, including RN #2, acknowledged that the portable oxygen concentrator should have been refilled regularly to prevent such occurrences. Resident #2, diagnosed with chronic respiratory failure and COPD, was found to be receiving oxygen at 2 LPM, contrary to the physician's order of 3 LPM. This discrepancy was observed on two separate occasions, and RN #1 confirmed the error, adjusting the oxygen flow to the correct level. The incorrect oxygen administration could have led to confusion and hypoxia, as noted by the staff. Resident #7, with a history of respiratory arrest and COPD, was using oxygen without a physician's order. The resident's care plan included continuous oxygen at 10 LPM, but the oxygen concentrator was set at 3 LPM. LPN #1 confirmed the absence of a physician's order for the oxygen therapy. The assistant director of nursing emphasized the importance of having a physician's order for oxygen use and ensuring that oxygen is administered as prescribed to prevent adverse outcomes.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Missing Oxygen Order and Improper Nebulizer Storage
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Missing Oxygen Order and Improper Nebulizer Storage: A resident receiving oxygen for sleep apnea had no physician order specifying the oxygen delivery rate, even though staff were setting the rate. In a separate observation, a resident’s nebulizer mask was left unbagged on the nightstand while not in use. An LPN and the DON both stated respiratory equipment should be stored in a bag when not in use, and the facility policy required bagged storage between uses.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory equipment was not maintained and stored per standards
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory care was not provided in accordance with standards for several residents receiving oxygen and nebulizer treatments. Staff observed outdated oxygen tubing and humidification equipment, a nebulizer mouthpiece left out with tubing touching the floor, and oxygen supplies not stored properly. One resident’s oxygen was running at a higher flow than ordered, and staff confirmed the tubing and humidification items should be changed weekly and documented.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage and Dating of Oxygen and CPAP Equipment
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Improper Storage and Dating of Oxygen and CPAP Equipment: Two residents had oxygen equipment and CPAP items observed out of proper storage, including nasal cannula tubing on the floor, undated or empty humidifiers, and a CPAP mask on the floor. One resident had OSA with an order for oxygen at bedtime and as needed, and the other had COPD with orders for nightly CPAP and continuous oxygen. Staff stated the tubing, humidifier, and CPAP mask should be stored and dated per facility practice, and the facility policy required weekly changes and dating of oxygen equipment.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Contaminated oxygen tubing was placed back on a resident
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident receiving continuous O2 via nasal cannula had the cannula found on the floor while the concentrator was running. A CNA placed the cannula back on the resident’s face before an LVN identified it as contaminated and replaced the cannula and tubing. The resident had diagnoses including hypertensive heart disease and atherosclerotic heart disease, and the care plan called for O2 at 2 L/min to maintain O2 sats above 92%.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Stored Sanitarily
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD, a respiratory infection, continuous O2, and CPAP therapy had CPAP and nasal cannula tubing observed lying on the bed and the CPAP mask draped over the machine instead of being stored in a sanitary container. Staff interviews showed inconsistent understanding of how to store the respiratory equipment, and the facility did not provide the requested storage of the equipment when not in use.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Properly Labeled or Stored
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory equipment was not properly changed, labeled, or stored for a resident receiving nebulizer treatment. An undated aerosol mask attached to a nebulizer was observed on the resident’s nightstand, open to air, and remained there on a later observation. An LPN confirmed the mask was open to air and not stored properly, and the DON confirmed the mask and tubing had not been dated or stored properly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Colorado

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Colorado — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release July 29, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙