F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
E

Failure to Timely Change Respiratory Equipment

Arbor Hills Rehabilitation And Healthcare CenterEagle Lake, Texas Survey Completed on 06-13-2024

Summary

The facility failed to provide timely respiratory care for a resident, specifically in changing respiratory equipment, which is crucial for infection control. The resident, an elderly female with a history of cerebral infarction, heart failure, and pneumonia, among other conditions, was observed with respiratory equipment that had not been changed since 05/20/24, despite the facility's policy requiring weekly changes. This included a breathing mask and an oral yankauer suction tip, both of which were dated over three weeks prior to the observation. Interviews with facility staff revealed a lack of awareness and adherence to the facility's infection control policies. The Infection Control Nurse/ADON stated that respiratory equipment should be changed weekly, but the LVN responsible for administering breathing treatments was unsure of the frequency, estimating changes to occur once or twice every two weeks. The DON confirmed the weekly change policy but was uncertain if nonclinical staff had been adequately trained on their responsibilities as ambassadors, who are supposed to check on residents and ensure equipment is properly dated and maintained. The ambassador assigned to the resident admitted to checking for dates on equipment but was not familiar with the specific requirements for changing respiratory equipment, as she was not a clinical staff member. This lack of knowledge and oversight contributed to the failure to maintain proper infection control practices, potentially placing the resident at risk for respiratory infections.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Missing Oxygen Order and Improper Nebulizer Storage
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Missing Oxygen Order and Improper Nebulizer Storage: A resident receiving oxygen for sleep apnea had no physician order specifying the oxygen delivery rate, even though staff were setting the rate. In a separate observation, a resident’s nebulizer mask was left unbagged on the nightstand while not in use. An LPN and the DON both stated respiratory equipment should be stored in a bag when not in use, and the facility policy required bagged storage between uses.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory equipment was not maintained and stored per standards
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory care was not provided in accordance with standards for several residents receiving oxygen and nebulizer treatments. Staff observed outdated oxygen tubing and humidification equipment, a nebulizer mouthpiece left out with tubing touching the floor, and oxygen supplies not stored properly. One resident’s oxygen was running at a higher flow than ordered, and staff confirmed the tubing and humidification items should be changed weekly and documented.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage and Dating of Oxygen and CPAP Equipment
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Improper Storage and Dating of Oxygen and CPAP Equipment: Two residents had oxygen equipment and CPAP items observed out of proper storage, including nasal cannula tubing on the floor, undated or empty humidifiers, and a CPAP mask on the floor. One resident had OSA with an order for oxygen at bedtime and as needed, and the other had COPD with orders for nightly CPAP and continuous oxygen. Staff stated the tubing, humidifier, and CPAP mask should be stored and dated per facility practice, and the facility policy required weekly changes and dating of oxygen equipment.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Contaminated oxygen tubing was placed back on a resident
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident receiving continuous O2 via nasal cannula had the cannula found on the floor while the concentrator was running. A CNA placed the cannula back on the resident’s face before an LVN identified it as contaminated and replaced the cannula and tubing. The resident had diagnoses including hypertensive heart disease and atherosclerotic heart disease, and the care plan called for O2 at 2 L/min to maintain O2 sats above 92%.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Stored Sanitarily
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD, a respiratory infection, continuous O2, and CPAP therapy had CPAP and nasal cannula tubing observed lying on the bed and the CPAP mask draped over the machine instead of being stored in a sanitary container. Staff interviews showed inconsistent understanding of how to store the respiratory equipment, and the facility did not provide the requested storage of the equipment when not in use.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Equipment Not Properly Labeled or Stored
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory equipment was not properly changed, labeled, or stored for a resident receiving nebulizer treatment. An undated aerosol mask attached to a nebulizer was observed on the resident’s nightstand, open to air, and remained there on a later observation. An LPN confirmed the mask was open to air and not stored properly, and the DON confirmed the mask and tubing had not been dated or stored properly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Texas

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Texas — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release July 29, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙