F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
E

Deficient Call Light Systems in LTC Facility

Alden Lakeland Rehab & HccChicago, Illinois Survey Completed on 07-11-2024

Summary

The facility failed to ensure that the call light systems were operating in good working condition, affecting five residents. Resident R143 reported that their call light was not functioning, causing anxiety and necessitating the use of a personal cell phone to call for assistance. The surveyor observed that the call light system in R143's room was not activated when the string was pulled, and the Registered Nurse confirmed the malfunction. No alternative device was provided to R143 to signal for help while the call light was in disrepair. Residents R61, R114, and R31 also experienced issues with their call light system. R61 stated that staff did not respond when the call string was pulled, and the surveyor observed that the light outside their room did not illuminate when activated. The Licensed Practical Nurse confirmed the issue and acknowledged that staff might miss the call light, leading to a lack of assistance for the residents. No alternative call system was provided to these residents. Resident R67 was observed with a call light string that was not connected to the switch, rendering the system inoperative. The Certified Nursing Assistant confirmed the disconnection and noted that other call lights on the unit were also not working. No alternative call system was provided to R67. The Director of Nursing affirmed that all residents should have a working call light and that bells should be provided when call lights are not functioning.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0919 citations
Bathroom Emergency Call Light Pull Cords Not Accessible
E
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Bathroom emergency call light pull cords in several resident rooms were observed hanging above the handrail and not readily accessible from the floor. The Maintenance Director stated they were unaware of any regulation related to bathroom pull cords, and the Administrator stated the expectation was that resident bathroom pull cords should be accessible if a resident was on the floor.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Nonfunctioning Resident Call Lights
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Nonfunctioning Resident Call Lights: A resident reported waiting a long time for staff to answer her call light and said it had been disconnected from the wall extender, while another resident said her call light had been unplugged and she had to get help by having a friend use his call light. Staff and residents also reported repeated call light problems for another resident, and the DON stated call lights were expected to always be in working order; the facility did not have a call light policy.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Nonfunctioning Resident Call Buttons
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Nonfunctioning Resident Call Buttons: Two residents had call buttons that did not work, and testing showed no light or alarm at the nurse’s station. An LVN later gave each resident a handbell as a temporary replacement, but one resident reported ringing it overnight without response and woke up with a soaked brief. Staff interviews showed the handbells were not clearly communicated to all aides, and the MAINTD said the call system problem was related to cords wrapped around bed frames causing a short.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Call Light Not Accessible in Shower Area
E
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

A resident with hemiplegia/hemiparesis, HF, gait impairment, and muscle weakness required max assist with showering and toilet hygiene, and the care plan included keeping the call light within reach. During observation, the resident was seated in a shower chair while an NA left the room, and the bathroom call light box was located across the room next to the toilet, making it unreachable from the shower. Staff and the DON stated residents were not to be left unattended in the shower, but the new call light setup left no accessible call light or cord in the shower area despite the facility policy requiring call lights at the bedside, toilet, and shower/bathing facility.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Nonfunctioning Call Light System on Cove Unit
E
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

A resident call light system on the Cove unit was not working, and multiple residents reported they could not summon help when needed. Residents said their call lights did not illuminate or sound, some had to yell, wheel into the hallway, or wait for staff rounds, and not all residents were given bells. Staff knew parts of the system were down, but communication and response were inconsistent, and maintenance was not notified immediately.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Call Light Not Available Within Residents’ Reach
D
F0919 F919: Make sure that a working call system is available in each resident's bathroom and bathing area.
Short Summary

Call Light Not Within Reach for Two Residents: A facility failed to ensure a working call system was available in resident rooms and bathing areas. Two cognitively intact residents who needed assistance with ADLs reported they had no call light or bell to summon staff, and observations confirmed no call light or bell was present in their rooms. The facility policy required call lights to be within the resident’s reach at all times.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across Illinois

Get a heads-up on the newest immediate-jeopardy (J–L) citations in Illinois — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release July 29, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙