F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
E

Failure to Inspect Bed Frames, Mattresses, and Bed Rails for Entrapment Risks

Brunswick Health Care CenterBrunswick, Missouri Survey Completed on 04-18-2024

Summary

The facility failed to complete inspections of bed frames, mattresses, and bed rails as part of their regular maintenance program to identify areas of possible entrapment for six residents. This deficiency was identified through observation, interview, and record review. The facility's policy on bed rail safety checks was not adhered to, as there was no evidence of inspections being conducted for the residents' beds, which included bariatric beds and beds with assist bars or side rails. The residents involved had various levels of cognitive and physical impairments, requiring substantial assistance with bed mobility and transfers, and some had physician orders for bed rails to assist with mobility and positioning. Resident #2, who was cognitively intact but required maximal assistance with bed mobility, was observed in a bariatric bed with an assist bar. However, there was no documentation of an inspection of the bed frame, mattress, or assist bars. Similarly, Resident #31, who had impaired ADL functioning and required substantial assistance with bed mobility, was observed with assist bars on both sides of the bed, but no inspection records were found. Resident #12, who had hemiplegia following a stroke and used bilateral 1/2 side rails for bed mobility, also had no documented entrapment zone assessment or physician order for side rails prior to the survey. Other residents, including Resident #23 with severe cognitive impairment and obesity, Resident #15 with hemiplegia and Parkinson's disease, and Resident #30 with severe cognitive impairment and Alzheimer's, were also found to have bed rails without documented inspections. Observations showed these residents using bed rails for mobility and positioning, but there were no records of entrapment zone assessments or measurements. The Director of Nursing confirmed that no entrapment zone assessments or measurements had been completed prior to the annual survey, indicating a systemic failure in the facility's maintenance and safety protocols for bed rails and mattresses.

Penalty

5 days payment denial
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0909 citations in Ohio
Incompatible Mattress and Bed Frame Resulting in Unsafe Bed Setup
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

A resident with multiple medical conditions was found to have a low air loss mattress that overhung the bed frame by about five inches, preventing the installation of a grab bar on one side. The Maintenance Director confirmed the mattress was too large for the frame and not fully supported.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Incompatible Mattress and Bed Frame
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

A resident with cognitive impairment and multiple health conditions was found to have a mattress that did not fit the bed, creating a large gap. The facility lacked a program for regular inspections of bed frames, mattresses, and bedrails. The Maintenance Supervisor noted a nine-inch gap, and the DON confirmed the bed was a rental, acknowledging the need for an extender or longer mattress.

Fine: $60,320
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Unsafe Bed Rail and Mattress Configuration
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

A resident with quadriplegia was found to have unsafe gaps between their mattress and bed rails, with measurements of 5.25 inches on the left and 3.25 inches on the right. The DON confirmed these gaps were too large and potentially unsafe, and the facility could not provide manufacturer guidelines for acceptable distances.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Assess Side Rails/Enabler Bars
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

The facility failed to assess the use of side rails/enabler bars for two residents with cognitive impairments and multiple diagnoses. Interviews revealed that enabler bars were not assessed for proper fit or need upon admission or routinely thereafter, despite the facility's policy requiring such assessments.

Fine: $83,53840 days payment denial
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Mattress Fit on Bed Frame
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

A facility failed to ensure a resident's mattress fit properly on the bed frame, resulting in a 12-inch gap between the headboard and the mattress. The resident, who had multiple diagnoses and was dependent on staff for all ADLs, was at risk due to this safety issue. The facility's policy on mattress inspection and bed safety was not followed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

65.1% of Ohio facilities received at least one citation during their inspection in the last 12 months.Will yours be survey-ready?

Surveyors issued 55 serious citations across Ohio in the last 12 months. See exactly what they're citing.

Get ready for your next survey

See what surveyors are citing in Ohio and spot your risk areas before they do.

Monthly Citation Reports

Have you been cited for this tag?

Save hours drafting a compliant Plan of Correction — AI built on real approved POCs.

Plan of Correction Writer

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release May 27, 2026) and official state health department websites — never guesswork.

Trusted by long-term care providers and associations.

Allegria Senior Living logo
FHCA logo
WeCare Centers logo
Care Rehab logo
An unhandled error has occurred. Reload 🗙