F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
D

Oxygen Therapy Not Set Up or Maintained as Ordered

National City Post AcuteNational City, California Survey Completed on 06-22-2026

Summary

The facility failed to provide respiratory care services in accordance with professional standards of practice for two residents who were receiving oxygen therapy. One resident had diagnoses including COPD and chronic respiratory failure with hypoxia and was ordered to receive oxygen at 3 LPM at all times, with adjustments allowed between 2 and 5 LPM to maintain an oxygen saturation between 88% and 92%. During observation, the resident was seated in a wheelchair with a nasal cannula connected first to a white concentrator that showed a low battery indicator and later to a green oxygen tank. The resident stated she did not feel as if she were getting enough oxygen, and her oxygen saturation was measured at 89% before the nasal cannula tubing was repositioned from behind her back, after which the saturation increased to 90%. Interviews showed conflicting accounts about who had set up the oxygen equipment and who was responsible for applying or adjusting oxygen. A CNA stated she had connected the nasal cannula to the concentrator and wheeled the resident to the dining room, while a nurse stated staff should have checked that the concentrator was adequately charged before bringing the resident to the dining room. The respiratory therapist later stated she had switched the nasal cannula from the concentrator to the oxygen tank after noticing the low battery indicator. Other staff stated that only nurses or RTs were allowed to apply oxygen or change oxygen settings, and the facility did not provide a policy outlining staff responsibilities for oxygen use. A second resident with COPD had an order for oxygen at 2 LPM as needed, with a humidifier when oxygen was delivered at 4 LPM or higher. During observation, the resident was in bed with a nasal cannula in place, and the humidifier attached to the oxygen concentrator was observed to be empty, with only droplets of water in the tubing. The resident stated she wanted the humidifier because she had previously experienced dry nostrils and a nose bleed from oxygen use. The DSD observed the empty humidifier and stated it should have been replaced before becoming empty, while a nurse stated the oxygen setup should match the physician's order and that any discrepancy should be clarified with the physician.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Oxygen equipment not maintained per order
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident who required oxygen therapy for SOB had an oxygen bubbler that was overdue for replacement, despite a TAR directing weekly changes of oxygen supplies including the bubbler and tubing. Staff confirmed the bubbler should have been changed weekly, and the DON and staff educator stated nurses were expected to follow the provider order as written and that all disposable oxygen pieces were to be changed weekly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Bi-pap Therapy Not Ordered or Verified at Admission
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD, acute respiratory failure with hypercapnia, and oxygen dependence was admitted with prior records showing nightly Bi-pap use and specific settings, but the facility did not have a Bi-pap order or settings in the chart and the baseline care plan did not address oxygen or Bi-pap. The resident reported he went without Bi-pap for three nights because a connector piece was missing, while the hospice RN, admitting LPN, ADON, and DON each stated they were unaware of the machine or did not verify the equipment and orders at admission.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Respiratory Care and Tracheostomy Care Not Provided as Ordered
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Respiratory care was not provided as ordered for one resident receiving O2, as the NC tubing was not dated. Tracheostomy care was also not provided as ordered for another resident: heavy thick mucus was visible under the trach, the ATM mask was soiled with mucus, the ATM and nebulizer tubing were not dated or stored in a bag, and the suction catheter tubing end was not stored in a bag. An RN confirmed the findings.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Change Oxygen Equipment per Physician Orders
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

The facility failed to follow physician orders for oxygen equipment for two residents with COPD-related diagnoses. Observations showed each resident’s O2 tubing and water bottle remained dated and unchanged beyond the ordered weekly schedule, and the DON confirmed the equipment should have been changed per the orders.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Ordered Continuous Oxygen
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Failure to maintain ordered continuous O2: A resident with stroke and COPD was observed without his NC while a CNA waited for a portable O2 unit to be filled. The CNA then applied the NC but did not turn on the liquid portable O2, and later wheeled the resident to the nurse’s station to have the nurse set the unit to the ordered liter flow. The resident’s physician had ordered continuous O2 2L to 4L NC.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage of CPAP Mask and Oxygen Tubing
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Improper Storage of CPAP Mask and Oxygen Tubing: A resident with COPD, DM, dialysis dependence, and continuous O2 orders had his CPAP/NIV mask observed lying on the bedside table and his O2 tubing wrapped under the concentrator handle instead of being stored in a sanitary container or bag. Staff interviews confirmed the equipment should be stored in a bag when not in use, but it was not being kept that way during observation.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Citation search

Search every citation & Plan of Correction

Go to search
Citation watch

Track new serious citations across California

Get a heads-up on the newest immediate-jeopardy (J–L) citations in California — where surveyors are focused right now.

Free · about one email a month

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release October 8, 2026) and official state health department websites — never guesswork.

In your survey window? See what surveyors are citing.

The Survey-Prep Report maps your facility's risk from 12 months of CMS and state citation data — what's being cited around you and what to check first. $129 one-time.

Get the Survey-Prep Report
An unhandled error has occurred. Reload 🗙

Connection lost — reconnecting… We couldn't reconnect automatically. Please reload the page to continue.